Open original FDA record ↗Recall record
- FDA identifier
- Z-0946-2022
- Company in FDA record
- Sartorius Corporation
- FDA record creation date
- Apr 15, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Virus Counter is an automated laboratory instrument that enables rapid quantification of intact viruses in solution. This instrument is not a medical device. The virus quantification is used in process development and manufacturing of various biotechnology products such as vaccines, gene therapy and personalized immunotherapy. Device is a Class 1 laser flow cytometer.
Source facts
- Action
- Sartorius sent customer notification dated March 11, 2022 to all affected customers. Customers were informed that a field service personnel will be dispatch to update the labeling and provide an updated User Information manual (operator s manual) *. This work occurs at no cost to you regardless of scenario. You need only make the unit accessible for a few minutes for the representative to complete the upgrade.
- Initiated
- 2022-03-11
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0946-2022
- Recall status
- Open, Classified
- CDRH record ID
- 192629
- FDA record created
- 2022-04-15
- Recall event number
- 89818