Open original FDA record ↗Recall record
- FDA identifier
- Z-0947-2015
- Company in FDA record
- Phadia US Inc
- FDA classification date
- Jan 12, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mismatch between the sample ID and the test result reported for all subsequent sample racks in that run.
Source facts
- Action
- ThermoFisher sent an Urgent Medical Device Product Correction letter to all affected customers. The firm will inform customers of the specific sequence of errors and operator actions that can trigger the event via letter and phone, and will request that they do not respond to the Barcode Read Error and let the sample with the error eject normally from the instrument and be rerun after the barcode read error has been resolved. The firm will issue a mandatory Phadia 1000 Instrument Software (ISW) update that will correct the issue and make efforts to install it within 45 days from date of availability. Customers were asked to return the written response form via fax to: 1-888-243-5214 Attention: Customer Support, Phadia 1000 ISW Product Correction Response; EMAIL: Compliance-us.idd@thermofisher.com Email Subject Line: Phadia 1000 ISW Product Correction Response. Customers with questions were instructed to contact Phadia US Technical Support at 800-346-4364, option 2. For questions regarding this recall call269-492-1957.
- Initiated
- 2014-11-20
- Root cause
- Software design
- Terminated
- 2015-06-30
- Distribution
- Nationwide Distribution including AZ, CA, CO, CT, FL, GA, HI IL, IN, KS, MD, MI, MN, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, and WA.
- Recall number
- Z-0947-2015
- Recall status
- Terminated
- CDRH record ID
- 131895
- FDA classified
- 2015-01-12
- Recall event number
- 69872