Product may display clinical reports with the study date and time in GMT format, which may not be obvious to the end user, who may expect to see the local time in the report.
Source facts
Action
Product notification was issued on 12/29/2008, instructing users to identify affected products, and instructing users to refer to image overlays for local study date and time information. The firm is planning a field correction expected to be available in "Q1 2009". For additional information, contact Philips at 650-293-2624.
Initiated
2008-12-29
Root cause
Software design
Terminated
2010-11-29
Distribution
International - Italy, Denmark, Germany, Netherlands, Sweden, Spain, Portugal, Switzerland, Spain, Turkey, France, Saudi Arabia, Australia, Thailand, New Zealand, Japan, and Brazil.
Recall number
Z-0948-2009
Recall status
Terminated
CDRH record ID
76977
FDA classified
2009-01-30
Recall event number
50751