Open original FDA record ↗Recall record
- FDA identifier
- Z-0949-2017
- Company in FDA record
- Merit Medical Systems, Inc.
- FDA classification date
- Jan 5, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Recalling dilators included with the 6F PreludeEASE Hydrophilic Sheath Introducer because the dilator tips were manufactured with an undersized inner diameter and therefore will not allow passage of an 0.021 (0.53 mm) guide wire.
Source facts
- Action
- Merit Medical sent an Urgent Product Recall Notice dated December 1, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to identify any affected devices within their facility, quarantine them, and discontinue use. Ensure that all personnel to whom the devices were distributed are made aware of this field action. Consignees were asked to work with their Merit Sales Representative to arrange product return. Consignees with questions were instructed to contact their Merit Sales Representative or Merit Customer Service at response@merit.com or at 801-208-4381. For questions regarding this recall call 801-208-4623.
- Initiated
- 2016-12-01
- Root cause
- Process change control
- Terminated
- 2017-04-18
- Distribution
- Nationwide Distribution
- Recall number
- Z-0949-2017
- Recall status
- Terminated
- CDRH record ID
- 151906
- FDA classified
- 2017-01-05
- Recall event number
- 75999