Open original FDA record ↗Recall record
- FDA identifier
- Z-0950-2017
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Jan 18, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
During an infusion, a false Air-in-Line (AIL) alarm would cause the infusion to be interrupted.
Source facts
- Action
- Carefusion (BD) sent a Medical Device Safety Notification letter dated December 2, 2016, to all affected customers. The letter informed customers that BD has identified an issue with the Alaris Pump module. The letter informs the customers that the firm has received reports of Air-In-Line (AIL) alarms that have occurred when no air is observed in the line. Customers were informed of the issue and of the potential risk involved. The letter informs the customers of the actions for users and additional actions for biomedical engineering. Customers with questions regarding recall related questions were instructed to contact BD Support Center at (888) 562-6018, 7-4pm PT, M-F or email supportcenter@carefusion.com. Customers with adverse event reports are instructed to contact Customer Advocacy at (888) 812-3266, 24 hrs/day, 7days/wk. or email customerfeedback@carefusion.com. For questions regarding this recall call 858-617-2000.
- Initiated
- 2016-12-02
- Root cause
- Component change control
- Terminated
- 2018-07-24
- Distribution
- Nationwide Distribution
- Recall number
- Z-0950-2017
- Recall status
- Terminated
- CDRH record ID
- 150442
- FDA classified
- 2017-01-18
- Recall event number
- 75466