Reason stated in the source
Engineering analysis has determined that the magnetic switch in certain devices may stick in the closed position. When the magnetic switch becomes stuck in the closed position treatment of ventricular or atrial tachyarrhythmias is inhibited, while bradycardia pacing is unaffected.
Source facts
- Action
- Firm issued a Dr. Doctor letter dated 06/23/05. Letter informs physicians of the issue, clinical implications and recommendations. UPDATE: An Advisory Update, dated 08/01/05 is being issued to physicians to inform them of a new programmer software application upgrade and the programmer screen alerts. Additionally regulatory approval was given for a new switch component that will be incorporated into all Guidant CRT-D products.
- Initiated
- 2005-06-24
- Root cause
- Other
- Terminated
- 2006-12-31
- Distribution
- worldwide. Including United States, Virgin Islands and countries such as Argentina, Australia, Austria, Barbados, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Cyprus, Czech Republic, Denmark, Eygpt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Labanon, Luxembourg, Marshall Islands, Mexico, Netherlands, New Zealand, Northern Mariana Islands, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom and Venezuela.
- Recall number
- Z-0951-05
- Recall status
- Terminated
- CDRH record ID
- 40195
- FDA classified
- 2005-06-30
- Recall event number
- 32496