ACCELERATOR APS Centrifuge Module being recalled due to potential for micro-cracks to form over time in the metal centrifuge rotor buckets that may cause the centrifuge buckets to disconnect from the rotor in the centrifuge during centrifugation.
Source facts
Action
Notification letters sent to consignees on 09/09/08. Consignees provided instructions for inspecting the buckets for cracks and were asked to contact customer service reps for replacement buckets. Consignees also asked to respond via fax using Customer Reply form acknowledging receipt, understanding and completion of correction letter.
Initiated
2008-09-09
Root cause
Device Design
Terminated
2010-10-04
Distribution
Nationwide Distribution --- including states of TX, UT, and VA.
Recall number
Z-0951-2009
Recall status
Terminated
CDRH record ID
73522
FDA classified
2009-02-05
Recall event number
49474