Open original FDA record ↗Recall record
- FDA identifier
- Z-0951-2013
- Company in FDA record
- Diagnostica Stago, Inc.
- FDA classification date
- Mar 14, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Diagnostica Stago has confirmed that sometimes the delta OD measured can be changed leading to a lowered percentage result.
Source facts
- Action
- Diagnostica Stago Inc. sent e-mail notifications on 1/28/2013 to all consignees (laboratory supervisor). The written field safety notice, entitled Important Field Safety Notice, with its response form was attached. The letter identified the affected product and gave a description of the defect. Customers were asked to confirm all the results obtained in the abnormal range by either of the two methods provided. Users are to complete and return the enclosed response form as soon as possible to the Stago Hotline via fax, 973-644-9348. Questions should be directed to the Stago Hotline at 800-725-0607.
- Initiated
- 2013-01-28
- Root cause
- Software design (manufacturing process)
- Terminated
- 2013-07-19
- Distribution
- Worldwide Distribution - USA and Canada.
- Recall number
- Z-0951-2013
- Recall status
- Terminated
- CDRH record ID
- 115819
- FDA classified
- 2013-03-14
- Recall event number
- 64320