Skip to content

Public FDA record

Z-0951-2017: Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed t

Recall record Z-0951-2017 for Natus Manufacturing Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0951-2017
Company in FDA record
Natus Manufacturing Limited
FDA record creation date
Jan 6, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly.

Source facts

Action
Natus Manufacturing Limited sent an "Urgent Field Safety Notice" letter dated April, 2016, via email. The letter described the problem and the product involved in the recall. Informed the consignees that a repair kit to correct the issue is available for repair of affected units. Requested consignees to complete the attached form and Faxto 847-949-6031 or e-mail it to Natus Quality Programs. In turn the firm will send the repair kit. For questions contact Technical Service at 888-496-2887. For questions regarding this recall call 888-496-2887.
Initiated
2016-10-15
Root cause
Mixed-up of materials/components
Terminated
2019-06-24
Distribution
Worldwide Distribution - US including NJ, NC, TX, IL, VA, NY, FL, CA, MS, WA, IN, and MA. and Internationally to Canada and Costa Rica.
Recall number
Z-0951-2017
Recall status
Terminated
CDRH record ID
151447
FDA record created
2017-01-06
Recall event number
75846

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗