Open original FDA record ↗Recall record
- FDA identifier
- Z-0951-2017
- Company in FDA record
- Natus Manufacturing Limited
- FDA record creation date
- Jan 6, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly.
Source facts
- Action
- Natus Manufacturing Limited sent an "Urgent Field Safety Notice" letter dated April, 2016, via email. The letter described the problem and the product involved in the recall. Informed the consignees that a repair kit to correct the issue is available for repair of affected units. Requested consignees to complete the attached form and Faxto 847-949-6031 or e-mail it to Natus Quality Programs. In turn the firm will send the repair kit. For questions contact Technical Service at 888-496-2887. For questions regarding this recall call 888-496-2887.
- Initiated
- 2016-10-15
- Root cause
- Mixed-up of materials/components
- Terminated
- 2019-06-24
- Distribution
- Worldwide Distribution - US including NJ, NC, TX, IL, VA, NY, FL, CA, MS, WA, IN, and MA. and Internationally to Canada and Costa Rica.
- Recall number
- Z-0951-2017
- Recall status
- Terminated
- CDRH record ID
- 151447
- FDA record created
- 2017-01-06
- Recall event number
- 75846