Open original FDA record ↗Recall record
- FDA identifier
- Z-0952-2013
- Company in FDA record
- Extremity Medical LLC
- FDA classification date
- Mar 15, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected screw size based on the depth gauge reading will be about 2 mm longer than intended.
Source facts
- Action
- Extremity Medical sent an "URGENT: FIELD SAFETY NOTICE" dated January 10, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Response Form was included for customers to complete and return. Contact Customer Service at 1 (973) 588-8980 for questions regarding this notice. On April 10, 2013, Extremity sent notifications to their customers at the surgeon level. They were asked to complete the attached product field action response form.
- Initiated
- 2013-01-10
- Root cause
- Component change control
- Terminated
- 2013-07-18
- Distribution
- Worldwide Distribution-USA including the states of Florida, Pennsylvania, New Hampshire, Oklahoma, California, Maryland, Arizona, Ohio, North Carolina, Texas, Virginia, Illinois, New York and Tennessee, and the countries of Australia, Belgium, Israel, Netherlands and United Kingdom.
- Recall number
- Z-0952-2013
- Recall status
- Terminated
- CDRH record ID
- 115841
- FDA classified
- 2013-03-15
- Recall event number
- 64330