Skip to content

Public FDA record

Z-0952-2021: Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used

Recall record Z-0952-2021 for Topcon Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0952-2021
Company in FDA record
Topcon Medical Systems, Inc.
FDA record creation date
Feb 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record

Source facts

Action
Topcon issued Urgent Medical Device Correction letter on 12/18/20 via email. Letter states reason for recall, health risk and action to take: Topcon Healthcare will determine original patient demographics from the image database. Using the originally entered patient information Topcon Healthcare will reprocess the merged patient recordsand images and form a unique ID by combining the ID, first & last name and date of birth. (i.e. 12345/John Doe/19860120). Customer Care team will be contacting you by phone prior to the December 22nd database correction to further describe the process and help with any questions that you may have. If you have any concerns, or would speak to Topcon Healthcare prior to this call, please feel free to contact us at 866-922-6278 (option 2).
Initiated
2020-12-18
Root cause
Software design
Terminated
2022-10-19
Distribution
AZ, MO, NC, NY, PA, TX
Recall number
Z-0952-2021
Recall status
Terminated
CDRH record ID
185215
FDA record created
2021-02-04
Recall event number
87086

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗