Z-0952-2022: therascreen KRAS RGQ PCR Kit (24) REF 874052 (Japan IVD, not released in the USA)

Recall record Z-0952-2022 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0952-2022
Company in FDA record
Qiagen Sciences LLC
FDA record creation date
Apr 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.

Source facts

Action
Qiagen issued Urgent Medical Device Correction therascreen KRAS RGQ PCR Kit (24), REF 870021 letter on 2/25/22. Letter states reason for recall, health risk and action to take: This notice is only relevant for the analysis of samples originating from non-small cell lung cancer (NSCLC) tissue. For continued use of the therascreen KRAS RGQ PCR Kit (24), check the Flags/Warnings in the Sample Result table for automated interpretation of results. Sample results with one or more of the Flags/Warnings listed in this notice should be regarded as invalid with respect to the KRAS status and KRAS Mutation status. Setup a new PCR run to repeat the affected sample(s). To minimize Flags/Warnings for the controls and samples, strict compliance to the IFU guidance should be followed, including but not limited to: Correct reagent mixing is required, and must be ensured at each mixing step during the assay setup. Placement of the Rotor-Gene Q MDx according to Installation Procedures site requirements Forward this information to all individuals and departments within your organization who are using the therascreen KRAS RGQ PCR Kit (24), REF. 870021. If you are not the end user, please forward this notice to the product end user. Review this notice with your laboratory/medical director. Complete the Acknowledgement of Receipt Form attached to this letter by March 10, 2022 and email it to quality.communications@qiagen.com. QIAGEN is revising the Instructions for Use, as described in this notice, to reduce any risk resulting from the calling of a false positive or false negative G12C (12CYS) mutation result for NSCLC samples due to the Flags/Warnings listed in this notice and to emphasize strict compliance to the IFU guidance to minimize Flags/Warning. If you have any questions or concerns, pleas.se contact your local QIAGEN Technical Service Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
Initiated
2022-02-25
Root cause
Device Design
Distribution
AZ CA CT FL GA IL IN MA MN NC NY OK OR PA TN TX WV
Recall number
Z-0952-2022
Recall status
Open, Classified
CDRH record ID
192912
FDA record created
2022-04-19
Recall event number
89914

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗