Z-0952-2023: cobas infinity central lab, Material Number 07154003001

Recall record Z-0952-2023 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0952-2023
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA classification date
Jan 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information System (LIS) is created with an Internal ID rather than the External ID sent from the LIS. The erroneous event occurs under rare circumstances where the date in the sample's barcode does not match the order date sent from the LIS, and can lead to the order being mismatched to an another patient's sample ID rather than the subject patient's sample ID. The software bug impacts multiple cobas infinity central lab software (versions 2.5.x , 3.01.x, 3.02.x, and 3.03.x) and occurs when the afflicted software has been configured with specific pre-conditions.

Source facts

Action
On December 12, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were informed of the product issue. A Roche Diagnostics representative will contact you in the near future to assess your cobas infinity configuration and to determine if it meets all pre-conditions. If so, your Roche representative will work with you to implement a workaround immediately in addition to working with you to apply a more permanent solution via a software patch once available. Customers should consult with a physician or pathologist at your facility to determine clinical implications specific to your patients. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about this recall.
Initiated
2022-12-12
Root cause
Software design
Distribution
Domestic distribution nationwide. Foreign distribution pending.
Recall number
Z-0952-2023
Recall status
Open, Classified
CDRH record ID
197412
FDA classified
2023-01-12
Recall event number
91299

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗