Open original FDA record ↗Recall record
- FDA identifier
- Z-0952-2026
- Company in FDA record
- Rocket Medical Plc
- FDA classification date
- Dec 18, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Source facts
- Action
- On November 12, 2025, Urgent Medical Device Correction Letters were sent to customers. Actions to be taken: 1. Ensure a copy of this letter and the updated IFU are available to all users or potential users of this device. 2. Complete the FSN acknowledgement form and return via the email provided.
- Initiated
- 2025-11-07
- Root cause
- Labeling design
- Distribution
- Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
- Recall number
- Z-0952-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217397
- FDA classified
- 2025-12-18
- Recall event number
- 97994