Skip to content

Public FDA record

Z-0953-2013: Optetrak Logic Comprehensive Knee Prosthesis System, Proximal Tibial Spacer and screws, size 3.5, 8mm, Ca

Recall record Z-0953-2013 for Exactech, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0953-2013
Company in FDA record
Exactech, Inc.
FDA classification date
Mar 15, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A dimensional mismatch was identified which can potentially allow for interference between devices.

Source facts

Action
Exactech Inc. sent a *** Important *** Product Market Recall Notice dated January 30, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Customers were instructed to immediately cease distribution or use of these products. 2. Extend this information to your accounts that may have this product in their possession. 3. Verify if you have any of the subject Optetrak Logic Proxima Tibial Spacer (PTS) in the specified lots. Please complete the attached fax back form and return it to Exactech within the next 5 working days. Thank you for your prompt attention to this matter. For further questions please call (352) 377-1140.
Initiated
2013-02-01
Root cause
Device Design
Terminated
2014-09-22
Distribution
Nationwide Distribution including the states of FL, AL, LA, PA, VA, IL, OH, NY, NC, KS, TN, NJ, CO, GA, CA, MD, MN, IN, & Puerto Rico.
Recall number
Z-0953-2013
Recall status
Terminated
CDRH record ID
116190
FDA classified
2013-03-15
Recall event number
64465

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗