Skip to content

Public FDA record

Z-0953-2021: Ethicon Dermabond Advanced Topical Skin Adhesive-topical skin adhesive to hold closed easily approximated

Recall record Z-0953-2021 for Ethicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0953-2021
Company in FDA record
Ethicon, Inc.
FDA record creation date
Feb 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product distributed with a red tape as part of the sterile packaging which could impact the seal and result in a sterility breach

Source facts

Action
Ethicon issued Urgent Medical Device Removal Letter on 1/4/21 via UPS second day. Receipt of the letter will be confirmed via UPS tracking. Letter states reason for recall, health risk, action to take: 1.Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email it to Stericycle at 866-853-2807 or Ethicon5582@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 5. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. 6. Customers are required to return unused impacted Dermabond Advanced" Topical Skin Adhesive subject to this recall that are in their inventory immediately. To receive replacement reimbursement, customers must return product subject to this recall no later than April 9, 2021. Any non-affected product and any product returned after the date specified will not receive replacement. 7. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling 800-
Initiated
2021-01-04
Root cause
Packaging process control
Terminated
2023-11-29
Distribution
US Nationwide
Recall number
Z-0953-2021
Recall status
Terminated
CDRH record ID
185311
FDA record created
2021-02-04
Recall event number
87088

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗