Open original FDA record ↗Recall record
- FDA identifier
- Z-0953-2023
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- Jan 12, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Defect in sterile pouch seal, incomplete seal, of lighted Retractors, increases the risk of contamination, which can lead to post-operative infection.
Source facts
- Action
- On 11/17/22 customers were verbally notified to quarantine affected devices. On 12/08/22, recall notices were sent to customers asking them to take the following action: 1) Discontinue use of affected devices. 2) Return affected devices to the recalling firm. 3) Complete and return the customer confirmation form that can be emailed to neuro.quality@medtronic.com 3) The notice needs to be shared with all who need to be aware within your organization and any organization where the potentially affected devices have been transferred. Customers with questions are encouraged to call/email: Medtronic Technical Services at 1-888-826-5603/rs.navtechsupport@medtronic.com
- Initiated
- 2022-12-08
- Root cause
- Equipment maintenance
- Distribution
- US Nationwide distribution including in the states of OH, CA, FL, NJ, NY, MN, IN, VA, WA, MS, SC, MO, GA, PA, KS, TN, LA, TX, ME, MA, AL, NC, MD, NH, IL, AZ, MI, NV, MT, OK, SD, RI, CO, CT, NM, NE, DE, WI.
- Recall number
- Z-0953-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197481
- FDA classified
- 2023-01-12
- Recall event number
- 91328