Skip to content

Public FDA record

Z-0953-2023: RadiaLux Lighted Retractor (Pink), REF: 50-101-1

Recall record Z-0953-2023 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0953-2023
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Jan 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Defect in sterile pouch seal, incomplete seal, of lighted Retractors, increases the risk of contamination, which can lead to post-operative infection.

Source facts

Action
On 11/17/22 customers were verbally notified to quarantine affected devices. On 12/08/22, recall notices were sent to customers asking them to take the following action: 1) Discontinue use of affected devices. 2) Return affected devices to the recalling firm. 3) Complete and return the customer confirmation form that can be emailed to neuro.quality@medtronic.com 3) The notice needs to be shared with all who need to be aware within your organization and any organization where the potentially affected devices have been transferred. Customers with questions are encouraged to call/email: Medtronic Technical Services at 1-888-826-5603/rs.navtechsupport@medtronic.com
Initiated
2022-12-08
Root cause
Equipment maintenance
Distribution
US Nationwide distribution including in the states of OH, CA, FL, NJ, NY, MN, IN, VA, WA, MS, SC, MO, GA, PA, KS, TN, LA, TX, ME, MA, AL, NC, MD, NH, IL, AZ, MI, NV, MT, OK, SD, RI, CO, CT, NM, NE, DE, WI.
Recall number
Z-0953-2023
Recall status
Open, Classified
CDRH record ID
197481
FDA classified
2023-01-12
Recall event number
91328

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗