Open original FDA record ↗Recall record
- FDA identifier
- Z-0955-2015
- Company in FDA record
- GE Healthcare
- FDA classification date
- Jan 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A required quality control test was not performed during installation associated with the software of the Revolution CT scanner.
Source facts
- Action
- GE Healthcare Planned Action (s) to repair defect or to bring product into Compliance letter dated January 16, 2015 to GE Healthcare. 1. A GE Healthcare service representative will perform a noise specification and tolerance calibration to correct all affected systems via Field Action Instruction (FMI 25457). 2. A customer notification letter which includes a statement that GE Healthcare will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. GE Healthcare Surgery will implement this CAP by May 1, 2015. You may proceed or continue with implementation of the CAP (subject to the conditions noted above). Call Center phone numbers: United States: 800 437-1171, Japan: 0120-055-919 Information For other countries, please contact your local GE Healthcare Service Representative if you have any questions or concerns regarding this notification.
- Initiated
- 2014-11-12
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2015-04-29
- Distribution
- US Distribution to the states of CA, FL, IL, NY, UT and WA.
- Recall number
- Z-0955-2015
- Recall status
- Terminated
- CDRH record ID
- 132293
- FDA classified
- 2015-01-26
- Recall event number
- 70133