Z-0955-2015: The Revolution CT is a multi-slice (256 detector row) CT scanner consisting of a gantry, patient table, s

Recall record Z-0955-2015 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0955-2015
Company in FDA record
GE Healthcare
FDA classification date
Jan 26, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A required quality control test was not performed during installation associated with the software of the Revolution CT scanner.

Source facts

Action
GE Healthcare Planned Action (s) to repair defect or to bring product into Compliance letter dated January 16, 2015 to GE Healthcare. 1. A GE Healthcare service representative will perform a noise specification and tolerance calibration to correct all affected systems via Field Action Instruction (FMI 25457). 2. A customer notification letter which includes a statement that GE Healthcare will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. GE Healthcare Surgery will implement this CAP by May 1, 2015. You may proceed or continue with implementation of the CAP (subject to the conditions noted above). Call Center phone numbers: United States: 800 437-1171, Japan: 0120-055-919 Information For other countries, please contact your local GE Healthcare Service Representative if you have any questions or concerns regarding this notification.
Initiated
2014-11-12
Root cause
Radiation Control for Health and Safety Act
Terminated
2015-04-29
Distribution
US Distribution to the states of CA, FL, IL, NY, UT and WA.
Recall number
Z-0955-2015
Recall status
Terminated
CDRH record ID
132293
FDA classified
2015-01-26
Recall event number
70133

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗