False Positive Results: Product labeling states cross contamination is not an issue for high viral load titer samples, but the firm has determined that a potential sample contamination may not be detected at these levels and could result in false positive results being reported.
Source facts
Action
Consignees were sent an Urgen Medical Device Correction letter dated 5/14/07 which instructs them to define and establish a lower detection limit, to use UNG to prevent carryover whenever possible, and to confirm low positive results through an independent experiment. A revised device correction letter dated 6/15/07 was subsequently sent to each consignee updating the reason for recall and associated risks.
Initiated
2007-05-14
Root cause
Other
Terminated
2009-12-17
Distribution
Nationwide.
Recall number
Z-0956-2007
Recall status
Terminated
CDRH record ID
52324
FDA classified
2007-06-22
Recall event number
37948