- FDA identifier
- Z-0956-2023
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jan 13, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Mechanical connection between the tabletop and table base may be lost.
Source facts
- Action
- A customer letter dated 12/16/2022 was issued to affected consignees. Consignees are advised the following: -Before treatment is started, check if both transversal guide rails and the roller bearing at the end of each rail are in the correct position. -For the affected OR tables, an inspection including installation of a safety measure will be carried out. Siemens is currently developing this mitigation. Once correction is available, the service department will contact the consignees to arrange a date to perform the corrective action. Consignees are requested to acknowledge the recall letter and return the acknowledgement form to Siemens promptly.
- Initiated
- 2022-12-16
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Austria, Belgium, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Italy, Japan, Kazakhstan, Kuwait, Lebanon, Monaco, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, and Vietnam.
- Recall number
- Z-0956-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197571
- FDA classified
- 2023-01-13
- Recall event number
- 91361