- FDA identifier
- Z-0956-2025
- Company in FDA record
- Murata Vios, Inc.
- FDA classification date
- Jan 16, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
Source facts
- Action
- Murata Vios notified their consignees via email on 06/21/2021. The notice explained the issue and provided workflow guidance to prevent the problem from occurring while a software update was being created. The firm followed up with another email on 08/24/2021 informing the consignees that the software update was ready for deployment and to request scheduling an appointment for the upgrade.
- Initiated
- 2021-06-21
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution in the states of NY, LA and the country of India.
- Recall number
- Z-0956-2025
- Recall status
- Completed
- CDRH record ID
- 211934
- FDA classified
- 2025-01-16
- Recall event number
- 96078