Skip to content

Public FDA record

Z-0956-2025: muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm r

Recall record Z-0956-2025 for Murata Vios, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0956-2025
Company in FDA record
Murata Vios, Inc.
FDA classification date
Jan 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.

Source facts

Action
Murata Vios notified their consignees via email on 06/21/2021. The notice explained the issue and provided workflow guidance to prevent the problem from occurring while a software update was being created. The firm followed up with another email on 08/24/2021 informing the consignees that the software update was ready for deployment and to request scheduling an appointment for the upgrade.
Initiated
2021-06-21
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of NY, LA and the country of India.
Recall number
Z-0956-2025
Recall status
Completed
CDRH record ID
211934
FDA classified
2025-01-16
Recall event number
96078

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗