Z-0957-2023: MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography system.

Recall record Z-0957-2023 for Mobius Imaging, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0957-2023
Company in FDA record
Mobius Imaging, LLC
FDA classification date
Jan 17, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker received one report of an injury to an untrained user that occurred while the untrained user was moving an Airo TruCT unit in reverse. The firm will increase visibility of instructions for users to consider when transporting the unit in reverse by providing labels to be affixed onto the AIRO.

Source facts

Action
On December 7, 2022, the firm notified customers via Urgent Medical Device Correction letters. Customers were informed that the firm would add additional warning labels to the Airo to increase visibility of instructions for users to consider when transporting an Airo TruCT unit in reverse. A service representative will contact your facility once labels are available (anticipated February 2023) to schedule a time to affix the labeling to your Airo TruCT unit. In the interim, please ensure all Airo TruCT users and other relevant personnel in your facilities are informed of this issue. If the devices have been distributed to other organizations, inform Stryker so that they may inform personnel at the new locations appropriately. If you have any questions or concerns regarding this notification, please contact shirleytechnicalsupport@stryker.com or call Stryker Technical Support at 978 386 9619.
Initiated
2022-12-07
Root cause
Use error
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Chinese Taipei, Denmark, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Malaysia, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Poland, Russian Federation, Saudi Arabia, Spain, Switzerland, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Recall number
Z-0957-2023
Recall status
Open, Classified
CDRH record ID
197341
FDA classified
2023-01-17
Recall event number
91258

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗