Z-0958-2007: IMx Theophylline Reagent Pack (List 1A81-20) In Vitro test. This reagent system is composed of one bottle

Recall record Z-0958-2007 for Abbott Diagnostic International, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0958-2007
Company in FDA record
Abbott Diagnostic International, Ltd.
FDA classification date
Jul 3, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A decrease in mP values (calibration curve Net Polarization value) has been observed with this reagent lot at all calibrator levels and may produce the following results: Calibration Error Code 148 (Polarization Too Small), Calibration Error Code 172 (Calibration Not Stored or Calibration Not Accepted) Controls out of range (when using a stored calibration curve).

Source facts

Action
An April 11, 2007, Product Recall Letter was sent to all customers who received the affected lot. All customers who received the affected lot will be contacted. The following communication types were used: Product Recall Letter & Customer Reply Form. Letters were sent to customers by Federal Express Priority mail. The Customers were instructed the following: To discontinue use of and destroy IMx Theophylline reagent lot 41610QI00. If they had forwarded the lots listed in the Product Recall letter to other laboratories, they were instructed to inform them of this Product Recall and provide a copy of this letter to them. They also we instructed to retain a copy of the communication for their records
Initiated
2007-04-11
Root cause
Other
Terminated
2011-06-10
Distribution
Worldwide: CA, IL, MA, NY, OR, PA, VA, and WV and countries of Germany and Singapore
Recall number
Z-0958-2007
Recall status
Terminated
CDRH record ID
52441
FDA classified
2007-07-03
Recall event number
37957

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗