Z-0958-2026: InPen App, Model/CFN Number: MMT-8061 (Android Users)

Recall record Z-0958-2026 for Medtronic MiniMed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0958-2026
Company in FDA record
Medtronic MiniMed, Inc.
FDA classification date
Dec 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.

Source facts

Action
On 11/13/2025, correction notices were emailed to customers who were asked to update to version 8.0.1 or higher as soon as possible to resolve this issue. 1.Verify your app version: -On the home screen, navigate to the Settings menu (lower right corner). - Select Help and Support. - Note the App Version displayed. 2. If you are running versions 8.0.0, update to the latest version 8.0.1 or higher via the Google Play Store: - Open the Google Play Store. - Tap your profile icon at the top of the screen, then select Manage apps & device. - Tap Updates available or search for the InPen app. - Tap Update next to the InPen app icon. Electronically acknowledge that you have read and understand this notification and have followed the actions listed For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option #1. The Customer Confirmation Form can also be emailed to the firm at rs.safetyinformation@medtronic.com
Initiated
2025-11-13
Root cause
Software Design Change
Distribution
US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom
Recall number
Z-0958-2026
Recall status
Open, Classified
CDRH record ID
217080
FDA classified
2025-12-18
Recall event number
97904

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗