Open original FDA record ↗Recall record
- FDA identifier
- Z-0959-2013
- Company in FDA record
- Allen Medical Systems, Inc.
- FDA classification date
- Mar 18, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall.
Source facts
- Action
- Allen Medical Systems sent a "Urgent Medical Device Recall" letter dated February 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken. Customers were instructed to: 1. Immediately remove from service your Allen Traction Boot with the serial number(s) XXXXX 2. Contact Allen Medical Customer Service at (800) 433-5774 for an RMA number and instructions to return your Allen Traction Boot for inspection and repair if necessary. There will be no cost to you. 3. Return your Allen Traction Boot to Allen Medical as soon as possible. In order to minimize the inconvenience, we will repair your Allen Traction Boot and return the device to you within 48 hours of receipt.
- Initiated
- 2013-02-28
- Root cause
- Process control
- Terminated
- 2013-08-07
- Distribution
- Worldwide Distribution- USA including the states of TX, NV, KY, GA, OH, PA, NY, AL, WA, IL and the countries of New Zealand and Japan.
- Recall number
- Z-0959-2013
- Recall status
- Terminated
- CDRH record ID
- 116368
- FDA classified
- 2013-03-18
- Recall event number
- 64509