Open original FDA record ↗Recall record
- FDA identifier
- Z-0959-2019
- Company in FDA record
- Shimadzu Medical Systems Usa Com
- FDA record creation date
- Feb 27, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure.
Source facts
- Action
- Shimadzu sent an Urgent Medical Device Correction Notice letter dated January 2, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. The letters to the end users explain the problem and a workaround. A service representative will contact the end user to schedule a software update.
- Initiated
- 2019-01-14
- Root cause
- Software design
- Terminated
- 2024-06-07
- Distribution
- US Nationwide Distribution in the states of IN, OH, IL, SC
- Recall number
- Z-0959-2019
- Recall status
- Terminated
- CDRH record ID
- 170523
- FDA record created
- 2019-02-27
- Recall event number
- 81981