Z-0959-2021: PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC,

Recall record Z-0959-2021 for Prytime Medical Devices, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0959-2021
Company in FDA record
Prytime Medical Devices, Inc.
FDA record creation date
Feb 5, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guidewire included in the kit will not pass through the access needle to facilitate insertion of an introducer sheath and could therefore cause procedure delay or injury.

Source facts

Action
On 12/14/2020, Prytime Medical sent an "Urgent Medical Device Recall Notification" to affected consignees via FedEx. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1) Please immediately review your inventory for the recalled lot specified in this notification. Quarantine and do NOT use this lot. 2) If you have an urgent need for this kit, and dont have an alternative lot available then please ensure that you have a backup 18 gauge access needle readily available at the point of use. Do not use the 21G access needle in the kit. 3) Immediately complete and return the attached Customer Response Form to the contact on the form. a. Please specify the number of affected kits in your possession. b. Prytime will promptly contact you to arrange return and replacement shortly after receiving the completed form. 4) Share this recall notification with any users of the product within your facilities, or anyone else you may have sent the product to, in order to ensure they are also aware of this recall. 5) Report any adverse health consequences experienced with the use of this product to Prytime using the attached form, or by email at recall@prytimemedical.com, or by phone at (210) 340-0116. If you have any questions or discover any additional REBOA Convenience Set KT1835C Lot MFG3590 at your facility please contact the Recall Coordinator by phone at (210) 340-0116, or by email at recall@prytimemedical.com.
Initiated
2020-12-14
Root cause
Mixed-up of materials/components
Terminated
2022-09-28
Distribution
US Nationwide distribution including in the states of NJ, NY, CA, TX, MI, MO, CO, MI, VA, DE, TN, MO, MT, AZ, FL, WI, SC, WA, TX, CT, IN, PA, ME, WI, MA, GA, OH, AR, UT, NY OUS: None
Recall number
Z-0959-2021
Recall status
Terminated
CDRH record ID
185333
FDA record created
2021-02-05
Recall event number
87099

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗