Z-0959-2022: 4.0mm x 12mm Cannulated Screw, Short Thread; Part number 3005-40012

Recall record Z-0959-2022 for Acumed LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0959-2022
Company in FDA record
Acumed LLC
FDA record creation date
Apr 20, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture, joint impingement or damage, or soft tissue damage.

Source facts

Action
Urgent Medical Device recall notice dated 3/8/22 was send out to customers via mail. Special Instructions: IMMEDIATELY stop using and/or distributing the product and complete the following actions 1) Perform a physical count of your inventory. 2) Remove the affected product, such that they are out of service. 3) Record this data on the survey and Return the survey via email to RecallNotification@acumed.net. 4) To facilitate return of product. a) Independent Sales Consultants: contact Agent Inventory via email at agentinventory@acumed.net b) Domestic Direct Customers: contact Business Services via email at businessservices@acumed.net or contact your agent directly to be assigned a Return Material Authorization number and applicable shipping information. c)International Customers: contact Business Services via email internationalbusinessservices@acumed.net or via phone at (888) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada) to be assigned a RMA number and applicable shipping information. 6) Once completed, please return the affected product using the Return Label. For patients implanted with these affected devices, firm advises physicians to monitor patients. The implanting and treating physicians are in the best position to exercise medical judgment for their patients and should make the final decision on indications for additional treatment depending on the patient s symptoms. Physicians are urged to notify of any adverse events or product safety issues. Please note that this recall extends to the user level. If you have distributed the affected product, please notify your consignees about this issue and ensure that any inventory is taken out of service. Please contact Business Services via email at BusinessServices@acumed.net or via phone at (888) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada) with any questions regarding this notification.
Initiated
2022-03-09
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, Puerto Rico and the country of South Africa.
Recall number
Z-0959-2022
Recall status
Completed
CDRH record ID
192460
FDA record created
2022-04-20
Recall event number
89771

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗