Z-0960-2025: VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit,

Recall record Z-0960-2025 for Percussionaire Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0960-2025
Company in FDA record
Percussionaire Corporation
FDA classification date
Jan 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

Source facts

Action
On 12/23/2024, the firm sent an "Urgent Medical Device Notification" to customers informing them that due to recently becoming aware through one customer complaint that the venturi component transiently stopped oscillating while in use. The patient experienced a slight desaturation (low oxygen). When the Phasitron is stuck, the system continues to deliver pressures and volumes, but with reduced amplitude. The set values return when the venturi resumes oscillation. Customers are instructed to: 1. Please review all inventory for impacted lots referenced in the table below. 2. An enhanced Phasitron pre-use check has been developed to evaluate circuits for this issue which replaces the standard pre-use verification until affected products have been replaced at your facility. Please post the instructions provided in Appendix 2 in all areas of your facility. This enhanced pre-use check is to be completed for each new Phasitron breathing circuit in the lots identified below. 3. If product fails the pre-use check found in Appendix 2, DO NOT USE and dispose per your institutional protocol and contact FSCA@sentec.com for product exchange. 4. If a product malfunction is identified during use, please IMMEDIATELY STOP USE of the device and dispose according to your institutional protocol and contact FSCA@sentec.com for product exchange. 5. Complete and Return Acknowledgement form (see Appendix 1) after reviewing and implementing the requested actions by January 31, 2024. 6. After receipt of replacement product: o Dispose remaining product from affected lots per facility protocol. For any question, email FSCA@sentec.com.
Initiated
2024-12-23
Root cause
Process control
Distribution
Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.
Recall number
Z-0960-2025
Recall status
Open, Classified
CDRH record ID
211696
FDA classified
2025-01-23
Recall event number
95979

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗