Skip to content

Public FDA record

Z-0961-2019: Alcon or Endure Ophthalmic Microscope Product Usage: Used for low magnification visualization during oph

Recall record Z-0961-2019 for Alcon Research Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0961-2019
Company in FDA record
Alcon Research Laboratories Ltd.
FDA classification date
Feb 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the optical head to detach from the stand due to the lack of thread adhesive.

Source facts

Action
On July 17, 2018 the firm sent letters to their consignees with the following instructions: To assist us in this Voluntary Medical Device Correction, please take the following steps: 1. Inspect your microscope prior to continued use per the instruction above. 2. If either the Functional or Visual tests have failed, stop use of the device immediately and notify Alcon at 1-800-832-7827. 3. Forward this information immediately to all departments within your organization who may be using the Alcon or Endure LuxOR E71 Ophthalmic Microscope and Alcon or Endure LuxOR E71 Q-Vue Ophthalmic Microscope. 4. If applicable, provide a copy of this notification to any other organization I location to which the Alcon or Endure LuxOR E71 Ophthalmic Microscope and Alcon or Endure LuxOR E71 Q-Vue Ophthalmic Microscope has been sold or moved. 5. Return the attached Response Form via fax or email to Alcon.
Initiated
2018-07-17
Root cause
Device Design
Terminated
2021-04-28
Distribution
Worldwide Distribution - US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, Ml, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, Puerto Rico and US Virgins Islands Australia, Austria, Bahamas, Belarus, Canada, Chile, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, Germany, Guatemala, India, Indonesia, Italy, Jamaica, Japan, Kenya, Lebanon, Mexico, Netherland Antilles, Norway, Peru, Philippines, Poland, Romania, Slovakia, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Ukraine, United Kingdom, and Vietnam
Recall number
Z-0961-2019
Recall status
Terminated
CDRH record ID
170577
FDA classified
2019-02-27
Recall event number
82024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗