Z-0962-2022: SIEMENS SOMATOM with software syngo CT VA30A_SP4 Models: SOMATOM go.Now 11061610 & 11061618 SOMATOM go

Recall record Z-0962-2022 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0962-2022
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Apr 20, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan

Source facts

Action
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to customers via CT010/22/S via email/mail on March 7, 2022. Letter states reason for recall, health risk and action to take: There are no workarounds for these issues. Customers are urged to install the remote update as soon as it is released. How will the corrective action be implemented and what is the efficiency of the corrective actions? Siemens has developed software update syngo CT VA30A_SP5 to ensure more stable scanning workflows, reduce the number of user notifications, and to further improve the image quality and reduce image artifacts. This update will provide workflow improvements, bug fixes for performance and stability problems observed in installed base. Corrective action will be provided free of charge and will be remotely distributed via Update Instruction CT008/22/S or CT088/21/S (X.cite only). The remote software update process will require approximately 180 minutes for completion. If you have any technical or scheduling questions, please contact our service organization 1-800-888-7436. You may also reach to the Regional Unit Contact listed on the first page of this notification. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understood the content therein. Please include your Siemens system names, material numbers, serial numbers, and Siemens Functional Location numbers and email this information to: recallsandrefusals.team@siemens-healthineers.com. Additionally, a software solution (VA30A_SP5) to eliminate the root cause of this problem is being distributed via CT008/22/S or CT088/21/S (X.cite only) to all affected customers free of charge. Fol
Initiated
2022-03-07
Root cause
Software design
Distribution
Nationwide
Recall number
Z-0962-2022
Recall status
Open, Classified
CDRH record ID
192851
FDA record created
2022-04-20
Recall event number
89901

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗