Open original FDA record ↗Recall record
- FDA identifier
- Z-0962-2026
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Dec 19, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
Source facts
- Action
- On 11/20/2025, correction notices were mailed and emailed to customers: 1) If a pocket fails with the error message of "Duplicate Address Detected," users with the proper permissions can leverage the "Recover Storage Space" functionality to eject the affected pocket and clear the error. Note: This does not resolve the Firmware v1.48 issue. This only clears the error message from the "Recover Storage Space" screen. Content stored within the affected CUBIE pocket will still need to be manually accessed once the CUBIE pocket is ejected. 2) Facilities should have comprehensive policies and procedures that outline alternative ways to access items, which may include removing from a different device or from pharmacy. 3. A redundant supply of all life sustaining medication should be kept in a nearby location to help mitigate the length of delay. 4. Firm recommends making a "code-box" or "crash cart" of medications that are readily available outside of an automated storage device in the event of an emergency. Facility policies and procedures should clearly identify when it is appropriate to access emergency medications. Refer to product user guide for additional information. 5. Keys can be used to manually access unavailable drawers. Keys must be readily available and retrievable to constitute risk mitigation. Hospital policies and procedures should be devised by the facility indicating location of keys along with how they are to be obtained and used in downtime scenarios. 6. Share and disseminate this notice within your facility network. 7. Complete and return the response form via email to BDRC21@bd.com Firm will install the firmware 1.49 upgrade either remotely or via a field service visit. Note: Firmware updates to main devices will update the Aux when connected to the main device. Firm contacts: Remediation Support: US-Pyxis-Field-Action-Remediation@bd.com Technical Support: bd.com/self-service, Phone:1-800-727-6102
- Initiated
- 2025-11-20
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution including in the states of CO, IN, OH, TX, KS, FL, GA, NC, CA, HI, IL, NY, MI, VA, AL, LA, NE, KY, OK, AR, TN, WY, NJ, WI, SD, AZ, NV, MO, MA, ID, WV, DE, PA, WA, MT, CT, IA, MD, PR, MN, NM, OR, VT, MS, ME, AK, UT, DC, GU, SC, NH, RI, ND, AS and the countries of CH, CA, KW, QA, NL, GB, UY, AU, NZ, BH, TH, SG, CN, FR, LB, BS, HK, TR, PL, AR, BR, CL, CO, MX, ES, BM, SA, IL.
- Recall number
- Z-0962-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216381
- FDA classified
- 2025-12-19
- Recall event number
- 97712