Skip to content

Public FDA record

Z-0964-2011: S5 C5-System, Part No.: 58-00-00, Sorin Group Deutschland Gmbh, Lindberghstrasse, 25, Munchen, Germany, S

Recall record Z-0964-2011 for Sorin Group USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0964-2011
Company in FDA record
Sorin Group USA, Inc.
FDA classification date
Jan 26, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Touch screen may become unresponsive, inhibiting user input.

Source facts

Action
All affected US customers and distributors were notified by certified mail on Nov 12, 2010 via a Field Correction letter. They were told that they would be contacted by Sorin to arrange an appointment to inspect and replace affected touch screens on site. They were told that they could safely continue using their S5 systems in accordance with the Operator's Manual and previously supplied instructions until the inspection and necessary replacement had been completed. A customer response form was also included for completion and return to the firm. Customers can contact the firm if they have any questions.
Initiated
2010-11-12
Root cause
Device Design
Terminated
2012-07-13
Distribution
Worldwide Distribution.
Recall number
Z-0964-2011
Recall status
Terminated
CDRH record ID
96792
FDA classified
2011-01-26
Recall event number
57609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗