Skip to content

Public FDA record

Z-0964-2026: da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

Recall record Z-0964-2026 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0964-2026
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Dec 19, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure

Source facts

Action
On 8/21/2024, correction notices were sent to customers who were informed of the following: 1) Share this notification with all personnel at your site who perform and support da Vinci procedures. 2) Inform affected personnel when the correction has been completed. 3) Complete and return the acknowledgement form via email to Recalls@intusurg.com 4) Retain a copy of this letter and the acknowledgement form for your files. 5) Firm is actively working on making updated motor module assemblies available. Once available, an Intuitive representative will schedule a visit to replace the affected motor module sub-assembly. If you need further information or support concerning this notification, please contact the firm's Customer Service: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com
Initiated
2024-08-21
Root cause
Device Design
Distribution
US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.
Recall number
Z-0964-2026
Recall status
Open, Classified
CDRH record ID
217489
FDA classified
2025-12-19
Recall event number
98048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗