Three lots of MUM-1 may not meet their labeled shelf life, as a drop in staining intensity over time has been observed.
Source facts
Action
Biocare sent a Medical Device Recall notifications letter to all customers by traceable method. The letters identified the affected product, problem and actions to be taken. Users are asked to immediately discontinue used of the affected lots and dispose of them according to applicable national, state and local laws. Users are requested to complete and return the response form so that proper credit can be issues
800-799-9499.
Initiated
2014-12-12
Root cause
Incorrect or no expiration date
Terminated
2015-03-30
Distribution
Worldwide Distrinution - US Nationwide in the states of TX, WI, MO, WV, SC, MA, WA, TN, AZ, CA, IL, NY, LA and countries of Kuwait, India, Venezuela, New Zealand, Turkey, Uruguay.
Recall number
Z-0965-2015
Recall status
Terminated
CDRH record ID
132108
FDA classified
2015-01-13
Recall event number
70010