Skip to content

Public FDA record

Z-0965-2026: LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 170

Recall record Z-0965-2026 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0965-2026
Company in FDA record
ICU Medical, Inc.
FDA classification date
Dec 19, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.

Source facts

Action
Firm began notifying customers on 11/24/2025 via Urgent Medical Device Correction letters. Customers were instructed to ensure all users or potential users of LifeShield ISSS DLM 2.2.1 with Plum Duo 1.1 drug libraries are made aware of the issue and the following proposed mitigations: Until a software update is released, when performing the first Plum Duo v1.1 drug library review after the migration, the pharmacist is able to review the changes using alternate review mechanisms, as noted in the LifeShield Drug Library Management User Manual: a) Accessing the version history view of the drug library b) Review drug library reports in CSV or HTML format c) Assess each change within LifeShield DLM to verify the accuracy of the changes (i.e., view impacted medication ruleset hard dose rate limit or view a CCA to verify a maximum rate setting) The firm is developing a software update to correct the issue and will notify customers of the timing of the software update.
Initiated
2025-11-24
Root cause
Software Design Change
Distribution
US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.
Recall number
Z-0965-2026
Recall status
Open, Classified
CDRH record ID
217406
FDA classified
2025-12-19
Recall event number
98000

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗