Z-0966-2022: Puritan Bennett 980 Series Ventilator

Recall record Z-0966-2022 for Medtronic formerly Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0966-2022
Company in FDA record
Medtronic formerly Covidien
FDA classification date
Apr 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.

Source facts

Action
The firm notified consignees starting on March 25, 2022 via customer letter. Customers were informed of the issue and advised to institute workflow modifications to ensure visibility of the GUI visual alarm banner to ensure awareness of an active alarm condition in the absence of a functioning audible alarm and/or omni-directional LED visual alarm. If the audible alarm or omni-directional LED alarm fail, the clinical team should transfer the patient to another form of ventilation. The firm will correct the issue in a future software update. For questions contact your Medtronic representative or Technical Service at 1-800-255-6774 The firm sent additional correction notices to consignees that received affected devices before the software correction was distributed.
Initiated
2022-03-25
Root cause
Under Investigation by firm
Distribution
US Nationwide - Worldwide .Distribution
Recall number
Z-0966-2022
Recall status
Open, Classified
CDRH record ID
192691
FDA classified
2022-04-25
Recall event number
89820

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗