Open original FDA record ↗Recall record
- FDA identifier
- Z-0966-2026
- Company in FDA record
- ICU Medical, Inc.
- FDA classification date
- Dec 19, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
Source facts
- Action
- On November 24, 2025, the firm notified customers via Urgent Medical Device Correction letters. Consignees were asked to take the following actions: 1. Ensure all users or potential users of LifeShield DLM 2.2.1 are immediately made aware of this notification and proposed mitigations. 2. Until a software update is released, limit any newly created or edited Concentration Limits to one decimal place of precision for limits less than 100. ICU Medical is developing a software update to correct this issue and will notify customers of the timing of the software update.
- Initiated
- 2025-11-24
- Root cause
- Software Design Change
- Distribution
- US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.
- Recall number
- Z-0966-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217444
- FDA classified
- 2025-12-19
- Recall event number
- 98022