Six lots of U-Clip Removal Tools, catalog REM1860FW were incorrectly assembled. The nose cone attachment to the body of the removal tool does not contain sufficient adhesive which can cause the nose to separate from the body of the device.
Source facts
Action
Customers to be notified via a letter on 2/8/07. The letter will explain the situation and requests that the impacted product be returned to Medtronic for disposal. The customer will be requested to complete a "Recall Certificate" acknowledging the receipt of this information.
Initiated
2007-02-08
Root cause
Other
Terminated
2008-02-24
Distribution
Nationwide and Canada.
Recall number
Z-0968-2007
Recall status
Terminated
CDRH record ID
50304
FDA classified
2007-06-27
Recall event number
37300