Open original FDA record ↗Recall record
- FDA identifier
- Z-0968-2008
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Feb 7, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Fracture of Stem Neck - The firm recently performed laboratory testing indicates that there may be a potential increased risk of stem neck fractures when a V40 +16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese.
Source facts
- Action
- Stryker Orthopaedics sent an Urgent Product Recall notification letters on November 9, 2007, by Federal Express with return receipt. The letter informed users of potential increased risk of stem neck fracture, when a V40 + 16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese. The firm has asked that inventory be examined, removed and held in quarantine. A Stryker representative will contact the medical facilities to retrieve the product. If a representative has not contacted the facility to date, it is suggested that they contact Stryker at (201) 831-5972.
- Initiated
- 2007-11-02
- Root cause
- Other
- Terminated
- 2010-08-30
- Distribution
- Nationwide Distribution
- Recall number
- Z-0968-2008
- Recall status
- Terminated
- CDRH record ID
- 68197
- FDA classified
- 2008-02-07
- Recall event number
- 45902