The ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.
Source facts
Action
On July 24, 2017 an Urgent notice was issued instructing customers to place a copy a notice in all kits remaining in stock. This notifies customers that they should amend the claims for open stability in the instructions for use, to state use immediately after the 30 minute reconstitution period. Questions or concerns can be emailed to technical.services@randox.com
Initiated
2017-08-11
Root cause
Device Design
Terminated
2019-03-29
Distribution
Distribution US nationwide.
Recall number
Z-0969-2018
Recall status
Terminated
CDRH record ID
161282
FDA record created
2018-03-12
Recall event number
79076