Open original FDA record ↗Recall record
- FDA identifier
- Z-0970-2007
- Company in FDA record
- Fisher & Paykel Healthcare Inc
- FDA classification date
- Jun 27, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Plastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and there is an extremely remote possibility that the broken tab may then enter the patient's nasal cavity or mouth. Masks manufactured since April 2006 do not feat
Source facts
- Action
- Fisher & Paykel Healthcare Inc. commenced sending out letters by mail to consignees the week commencing 4 December 2006 advising consignees to quarantine and destroy affected product. A faxback form was also be provided for consignees to advise Fisher & Paykel Healthcare of quarantined and destroyed stock. The firm will supply consignees with replacement product as determined from completed faxback information. The firm will collate responses from consignees and will follow up with a letter to consignees who do not respond within a month of initial mail-out. The firm will then make telephone contact with any consignees who do not respond within a month of the second mail-out. Fisher & Paykel Healthcare Inc. estimated the completion of this voluntary recall by April 2007.
- Initiated
- 2006-11-24
- Root cause
- Other
- Terminated
- 2012-03-26
- Distribution
- Nationwide
- Recall number
- Z-0970-2007
- Recall status
- Terminated
- CDRH record ID
- 49660
- FDA classified
- 2007-06-27
- Recall event number
- 36955