Difficulty locking EXp tibial inserts into tibial trays, caused by a manufacturing condition where a lead in radius on the anterior locking tab is not present on certain insert lots. The absence of this lead in may prevent or make assembly of the insert and tray difficult.
Source facts
Action
On December 1, 2014, phone calls were made to all direct accounts to notify them of the recall and to request immediate recovery and return of the SC3425 EXp PS Tibial Tray Inserts and SC3453 EXp CR-HF Inserts from their inventory.
Initiated
2014-12-01
Root cause
Process design
Terminated
2015-05-06
Distribution
TX, MD, IN, KS, CA, PA, GA, WA, NY, IL.
Recall number
Z-0970-2015
Recall status
Terminated
CDRH record ID
132223
FDA classified
2015-01-14
Recall event number
70082