Z-0970-2019: C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation

Recall record Z-0970-2019 for PENTAX of America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0970-2019
Company in FDA record
PENTAX of America Inc
FDA record creation date
Mar 1, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur.

Source facts

Action
On 10/18/18, the firm, Pentax Medical, started contacting US Affected Customers by telephone to inform them of the pending action and to advise they discontinue use and quarantine affected devices. On 12/03/18, "URGENT MEDICAL DEVICE REMOVAL" letters dated 11/16/18 were mailed via USPS Certified Mail to its customers. Customer response forms and Customer Return Material Shipment Labels were mailed with the removal notices. Customers were informed that affected products should not be used and should be returned to the firm. In addition, customers were asked to ensure that all potential users in their facilities are made aware of the removal notice and the recommended actions. Customers were asked to complete and return the Field Action Response Form via: Fax to PENTAX Recall Coordinator at 201-799-4063 (alternate 201-391-4189) or Email a pdf copy to customeradvisories@pentaxmedical.com, or mail with the returned affected product to:Pentax of America Inc,,303 Convention Way Suite 1, Redwood City, CA 94063, Attention: 2018-009-R Action. . Customers with additional questions were encouraged to call 650-318-5899 (8:30 AM - 5:00 PM, Monday - Friday, PST).
Initiated
2018-12-03
Root cause
Nonconforming Material/Component
Terminated
2020-08-21
Distribution
Worldwide Distribution: US (nationwide) to states of:: CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA;; and countries: Canada, Germany, and Netherlands.
Recall number
Z-0970-2019
Recall status
Terminated
CDRH record ID
169563
FDA record created
2019-03-01
Recall event number
81695

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗