Skip to content

Public FDA record

Z-0970-2020: Urethral Dilation Balloon Catheter with Open Tip , Global Product No. G17844

Recall record Z-0970-2020 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0970-2020
Company in FDA record
Cook Inc.
FDA record creation date
Feb 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Source facts

Action
On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
Initiated
2019-01-04
Root cause
Process control
Terminated
2020-05-29
Distribution
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Recall number
Z-0970-2020
Recall status
Terminated
CDRH record ID
170630
FDA record created
2020-02-04
Recall event number
82026

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗