Reason stated in the source
Ortho Clinical Diagnostics received numerous complaints of lower than expected results for control fluids and patient samples upon initial calibration or following a recalibration using VITROS Chemistry Products Calibrator Kit 24 and VITROS Chemistry Products mALB Reagent.
Source facts
- Action
- All consignees were sent "Important Product Correction Notification" communications (dated 10/27/11) beginning on 10/27/11 informing them of the issue, and instructing them to discontinue use of their remaining inventory of VITROS Chemistry Products Calibrator Kit 24, VITROS Chemistry Products mALB reagent, and VITROS Chemistry Products mALB Performance Verifiers. For questions, please contact OCD's Customer Technical Services at 1-800-421-3311.
- Initiated
- 2011-10-27
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-02-12
- Distribution
- Worldwide Distribution - USA, including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
- Recall number
- Z-0971-2012
- Recall status
- Terminated
- CDRH record ID
- 106485
- FDA classified
- 2012-02-07
- Recall event number
- 60794