Open original FDA record ↗Recall record
- FDA identifier
- Z-0971-2018
- Company in FDA record
- Baxter Healthcare Corporation
- FDA record creation date
- Mar 12, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.
Source facts
- Action
- An Urgent Medical Device Correction, dated October 6, 2017, was sent to customers. The letter described the problem, identified the affected device, discussed the hazard involved, and action to be taken. Customers were asked to locate the affected products at their facility. Customers should continue to follow the instructions in document 41394 Rev. A, PREVENTIVE MAINTENANCE, for their entire fleet of pumps. Baxter will inspect affected pumps. To arrange for return of devices that have failed the flow rate accuracy test, contact Baxter Healthcare Medina. If you are a dealer, wholesaler, or distributor/reseller that distributed any product to other facilities, please notify your customers of this Urgent Medical Device Correction. If there are additional questions, customers can contact their Baxter Sales Rep or Baxter Healthcare Medina at 800-356-3454.
- Initiated
- 2017-10-06
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-04-11
- Distribution
- Worldwide Distribution - US, including the District of Columbia, Puerto Rico, and Canada.
- Recall number
- Z-0971-2018
- Recall status
- Terminated
- CDRH record ID
- 161516
- FDA record created
- 2018-03-12
- Recall event number
- 79177