Open original FDA record ↗Recall record
- FDA identifier
- Z-0971-2024
- Company in FDA record
- Draegerwerk Ag & Co. Kgaa
- FDA classification date
- Feb 1, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.
Source facts
- Action
- Draeger US was notified of the recall by their corporate office in Germany on November 17, 2023. All 50 units imported into the US at the time had not been distributed to end-users and were under the direct control of Draeger, Inc. Therefore, the firm did not issue a recall communication to customers. As a corrective action, an Instructions for Use (IFU) supplement has been added to these affected devices. The supplement outlines additional instructions regarding battery operation and inspection. Future shipments of Atlan devices will have the IFU supplement included with the device shipment.
- Initiated
- 2023-11-17
- Root cause
- Use error
- Distribution
- No devices were distributed to end-users (customers) in the US. All devices are currently under the direct control of Draeger, Inc.
- Recall number
- Z-0971-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204697
- FDA classified
- 2024-02-01
- Recall event number
- 93544