Open original FDA record ↗Recall record
- FDA identifier
- Z-0972-2017
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Jan 9, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following detection of a fault can result in the failure of a module in the high-voltage generator. - For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with graphics card may, in rare cases, result in the loss of image display in the examination room.
Source facts
- Action
- Siemens mailed a Customer Safety Advisory Notice (CSAN) dated December 8, 2016, to all affected customers with Customer Safety Advisory Notice AX002/16/S. Additionally, Siemens has initiated a software update via Update Instruction AX001/16/S. Their Service Organization will contact each customer to schedule an appointment to perform the corrective action; however, customers can make contact for an earlier appointment. Customers with questions should call 1-888-826-9702. For questions regarding this recall call 610-448-6461.
- Initiated
- 2016-12-08
- Root cause
- Software in the Use Environment
- Terminated
- 2019-04-09
- Distribution
- Nationwide Distribution
- Recall number
- Z-0972-2017
- Recall status
- Terminated
- CDRH record ID
- 151943
- FDA record created
- 2017-01-09
- Recall event number
- 76016