Skip to content

Public FDA record

Z-0972-2024: NovoTHOR Whole Body Light Pod, Gen 3.0 (1) Product Code S2188 - Reg (2) Product Code S2189 - Reg PRX (3)

Recall record Z-0972-2024 for THOR Photomedicine Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0972-2024
Company in FDA record
THOR Photomedicine Ltd
FDA classification date
Feb 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or close the canopy

Source facts

Action
On November 20, 2023, the firm notified customers via IMPORTANT MEDICAL DEVICE CORRECTION letters sent through email. A service agent representing THOR will contact customers within the next eight weeks to schedule the fitting of replacement ball studs at no cost to the customer. Customers who continue to use the NovoTHOR should be aware that it may potentially cause harm should the failure occur while users/patients are lifting or lowering the canopy. Customers should warn all users of potential risk, ensure someone is present to lift and lower the canopy, ensure the canopy is lifted and lowered gently and slowly, and ensure someone is present in the room or immediately outside the room during use.
Initiated
2023-11-20
Root cause
Component design/selection
Distribution
Domestic distribution to the states of Arizona California Colorado Connecticut Florida Georgia Hawaii Idaho Indiana Maryland New Jersey New York Oregon Pennsylvania Texas Washington Wisconsin. Foreign distribution to Australia, Canada, Italy, Malta, Norway, Switzerland, UAE, and UK.
Recall number
Z-0972-2024
Recall status
Open, Classified
CDRH record ID
205315
FDA classified
2024-02-01
Recall event number
93723

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗