- FDA identifier
- Z-0972-2024
- Company in FDA record
- THOR Photomedicine Ltd
- FDA classification date
- Feb 1, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or close the canopy
Source facts
- Action
- On November 20, 2023, the firm notified customers via IMPORTANT MEDICAL DEVICE CORRECTION letters sent through email. A service agent representing THOR will contact customers within the next eight weeks to schedule the fitting of replacement ball studs at no cost to the customer. Customers who continue to use the NovoTHOR should be aware that it may potentially cause harm should the failure occur while users/patients are lifting or lowering the canopy. Customers should warn all users of potential risk, ensure someone is present to lift and lower the canopy, ensure the canopy is lifted and lowered gently and slowly, and ensure someone is present in the room or immediately outside the room during use.
- Initiated
- 2023-11-20
- Root cause
- Component design/selection
- Distribution
- Domestic distribution to the states of Arizona California Colorado Connecticut Florida Georgia Hawaii Idaho Indiana Maryland New Jersey New York Oregon Pennsylvania Texas Washington Wisconsin. Foreign distribution to Australia, Canada, Italy, Malta, Norway, Switzerland, UAE, and UK.
- Recall number
- Z-0972-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205315
- FDA classified
- 2024-02-01
- Recall event number
- 93723